A Background to Return Loss Measurement
Getting Started with Return Loss
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
39
Measuring the Power Transmitted Through
the DUT
Unterminate the DUT and connect the DUT to a Power Meter so that
the power that is transmitted through the DUT can be measured. This
power is called
E
il
Figure 34
Measuring the Power transmitted through the Device Under Test
You can use
E
il
to calculate the insertion loss, see
Insertion Loss of the DUT” on page 43
Calculating the Return Loss of the DUT
The system may be represented by the general diagram shown below:
The reflected power, measured by the instrument, (
P
), from the
component with the known reflection factor, is given by the sum of:
• the part of the power, reflected by the component, which is
transmitted through the coupler, and
• the reflections due to the measurement system.
M
il
measurement
patchcord
DUT
Power Meter
E
il
Agilent 81610A
Reflectivity
R
DUT
Reflection-free
Termination
R
M
ref
P
ref
Agilent 81610A
Reflectivity
P
in
P
r
P
src
t
1
k
1
k
2
t
2
Summary of Contents for 81610A
Page 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Page 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Page 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Page 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Page 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Page 116: ......