A Background to Return Loss Measurement
Getting Started with Return Loss
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
37
Measuring the Power Transmitted Through
the Reflection Reference
N O T E
Note you can only measure the Front Panel Delta if you use the a
Reflection Reference Cable.
Connect your a Reflection Reference Cable to a Power Meter and
measure the transmitted power,
E
ref
. You can use
E
ref
to calculate the
front panel delta, see
“Calculating the Front Panel Delta” on page 41
Figure 30
Measuring the Power transmitted through the Reflection Reference
Measuring the Power when there are No
Reflections
Next, terminate the mesurement patchcord so that there are no
reflections from the end. All the power measured by the sensor now, is
due to the non-ideal nature of the measurement system. This is our
termination parameter.
Figure 31
Measuring the Power with the Connector Terminated
This measured power for the termination parameter is called
P
T
.
Power Meter
E
ref
M
ref
P
ref
Reference Cable
Agilent 81610A
P
in
P
r
Reflection Reference
Reflection-free
Termination
= 0
M
ref
P
ref
Agilent 81610A
Reflectivity
P
in
measurement patchcord
Summary of Contents for 81610A
Page 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Page 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Page 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Page 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Page 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Page 116: ......