Making a Return Loss Measurement
Getting Started with Return Loss
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
33
– [
Meas
], the power currently measured by the Return loss
Module’s internal power sensor in dBm
Figure 26
The Calibration Parameters Screens - Return Loss Diode
2
Press [Next]. The
Calibration parameters
screen for the
Monitor
Diode
appears, as displayed in Figure 27. This screen shows current
value for the following quantities:
– [
Para
], the parasitic power value measured by the Return Loss
module’s monitor diode in dBm. This value is determined by the
termination calibration, or you can use the default setting held in
the factory calibration.
– [
Ref
], the power measured by the Return Loss module’s monitor
diode during the reflectance calibration in dBm