3-3
Eye-Diagram Analyzer Reference
Performing Eye-Diagram Measurements
Performing Eye-Diagram Measurements
To perform the automatic eye-diagram measurements, use the Measure menu.
With the exception of extinction ratio, these measurements must be per-
formed in eye mode.
Automatic Measurements
The Measure menu’s top two softkeys automatically start measurements:
• EXTINCT RATIO
• MEASURE EYE
• Use the EXTINCT RATIO softkey to automatically compute the extinction ratio
in eye or eyeline modes. This measurement is a ratio of the most prevalent log-
ical one level divided by the logical zero level over one bit interval. When mak-
ing extinction ratio measurements in eyeline mode, the number of samples
should be increased from the default of 1000 to approximately 20000. This in-
sures that a number of traces are evaluated to compute the extinction ratio.
Summary of Contents for 71501D
Page 1: ...Agilent 71501D Eye Diagram Analysis User s Guide...
Page 7: ...1 Getting Started...
Page 34: ...Getting Started Steps for Setting Up Eye Diagram Analysis...
Page 35: ...2 Application Tutorials...
Page 43: ...2 9 Application Tutorials Tutorial 2 Measure in Optical Power Units...
Page 48: ...Application Tutorials Tutorial 4 Measure Laser Turn on Delay Laser Overshoot and Turn On Delay...
Page 80: ...Application Tutorials Tutorial 11 Create a Horizontal Histogram...
Page 81: ...3 Eye Diagram Analyzer Reference...
Page 113: ...3 33 Eye Diagram Analyzer Reference Applying Mask Testing The Mask Setup Submenu...
Page 151: ...4 Programming Commands...
Page 163: ...5 Specifications and Characteristics...
Page 170: ...Specifications and Characteristics Declaration of Conformity Declaration of Conformity...