2-21
Application Tutorials
Tutorial 6: Test to Industry Standards
Stop on and Display Trace Errors
The eye-diagram analyzer has the ability to stop data acquisition when a mask
violation occurs. The number of traces or errors that stop this data acquisition
can be specified. In addition, if you are in eyeline mode, you can separately
display the traces that have caused an error.
8
Set the bit interval and delay by pressing:
Setup
,
BIT INTVL
and enter
1
,
DELAY
and enter
–3
The mask will be offset to the right side of the display.
9
Set the analyzer so testing will stop after two errors have occurred by pressing:
#Errors ON OFF
ON and enter
2
10
If #Traces is set to on, the eye-diagram analyzer stops sweeping when either
the error or trace limit is reached. Turn the number of traces function off by
pressing:
#Traces ON OFF
OFF
11
Turn error tracing on by pressing:
err trc ON OFF
ON
Any trace which violates the mask shows on the lower-half of the screen.
12
Reset and start the trace and error counters by pressing:
Summary of Contents for 71501D
Page 1: ...Agilent 71501D Eye Diagram Analysis User s Guide...
Page 7: ...1 Getting Started...
Page 34: ...Getting Started Steps for Setting Up Eye Diagram Analysis...
Page 35: ...2 Application Tutorials...
Page 43: ...2 9 Application Tutorials Tutorial 2 Measure in Optical Power Units...
Page 48: ...Application Tutorials Tutorial 4 Measure Laser Turn on Delay Laser Overshoot and Turn On Delay...
Page 80: ...Application Tutorials Tutorial 11 Create a Horizontal Histogram...
Page 81: ...3 Eye Diagram Analyzer Reference...
Page 113: ...3 33 Eye Diagram Analyzer Reference Applying Mask Testing The Mask Setup Submenu...
Page 151: ...4 Programming Commands...
Page 163: ...5 Specifications and Characteristics...
Page 170: ...Specifications and Characteristics Declaration of Conformity Declaration of Conformity...