Getting Started
Steps for Setting Up Eye-Diagram Analysis
.
Step 4. Set Up the Measurement Conditions
Select the Mode
1
To select the mode, press:
Setup
,
diagram
,
EYE
,
EYELINE
, or
PATTERN
Select Single or
Continuous
Sweeps
2
Press the left-side
Trg,Cal
softkey.
3
Select one of the following sweep states:
• Press
SINGLE
to select single sweeps. Each additional press of this softkey
triggers another sweep.
• Press
CONT STOP
CONT to select continuous sweeps.
• Press
CONT STOP STOP to stop the sweeps.
Select Trigger
Source
4
Press the left-side
Setup
softkey.
5
Press
CH2
.
• If the pattern generator’s
CLOCK OUT
signal is connected to the
RF INPUT
2
connector, press
CLK OUT
.
• If the pattern generator’s
TRIGGER OUT
signal is connected to the
RF IN-
PUT 2
connector, press
TRG OUT
.
Summary of Contents for 71501D
Page 1: ...Agilent 71501D Eye Diagram Analysis User s Guide...
Page 7: ...1 Getting Started...
Page 34: ...Getting Started Steps for Setting Up Eye Diagram Analysis...
Page 35: ...2 Application Tutorials...
Page 43: ...2 9 Application Tutorials Tutorial 2 Measure in Optical Power Units...
Page 48: ...Application Tutorials Tutorial 4 Measure Laser Turn on Delay Laser Overshoot and Turn On Delay...
Page 80: ...Application Tutorials Tutorial 11 Create a Horizontal Histogram...
Page 81: ...3 Eye Diagram Analyzer Reference...
Page 113: ...3 33 Eye Diagram Analyzer Reference Applying Mask Testing The Mask Setup Submenu...
Page 151: ...4 Programming Commands...
Page 163: ...5 Specifications and Characteristics...
Page 170: ...Specifications and Characteristics Declaration of Conformity Declaration of Conformity...