CI Maintenance
To Assemble the CI Source
7000/7010 Series TQ GC/MS Operating Manual
235
To Assemble the CI Source
Materials needed
•
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Hex ball driver, 2.0 mm (8710-1804)
•
Wrench, open-end, 10 mm (8710-2353)
Procedure
Refer to the exploded parts view and the CI source parts list while using this
procedure. (See
and
1
Assemble the ion focus lens, entrance lens, and lens insulator. (See
2
Slide the drawout plate and the drawout cylinder into the source body. (See
3
Slide the parts assembled in step 1 into the source body.
4
Install the setscrew that holds the lenses in place.
5
Attach the ceramic disk to the repeller and place on top of the source body.
6
Place the heater block assembly on top of the source body.
7
Reinstall the dummy filament, and the filament and attach with the
setscrews.
CAUTION
Always wear clean gloves when working in the analyzer chamber to avoid
contamination.