CI Maintenance
To Remove the EI HES Radiator
208
7000/7010 Series TQ GC/MS Operating Manual
To Remove the EI HES Radiator
Materials needed
•
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
•
Tweezers (8710-2460)
•
T10 Torx screwdriver (5182-3466)
Procedure
1
Vent the MS. (See
2
Open the front analyzer chamber. (See
3
Remove the EI HES source. (See
4
Disconnect the wires from the radiator or CSB. Do not bend the wires any
more than necessary. (See
“To Connect or Disconnect EI HES Radiator
5
Use a T10 Torx screwdriver to loosen the two screws that secure the radiator
to the analyzer and place the radiator in its storage container. (See
WARNING
The analyzers, GC/MS interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
CAUTION
Use an antistatic wrist strap, and take other antistatic precautions before
touching analyzer components.
CAUTION
When disconnecting leads, pull on the connectors, not on the wires.