General Maintenance
To Install the EI HES Filaments
194
7000/7010 Series TQ GC/MS Operating Manual
To Install the EI HES Filaments
Materials needed
•
Filament assembly, high efficiency dual filament (G7002-60001)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
•
T6 Torx driver, (8710-2548)
Procedure
1
Remove the old filament. (See
“To Remove the EI HES Filaments”
2
Insert the three pins on the dual filament through the rear of the filament
block. (See
3
Place the filament block in the source mount.
4
Use a T6 Torx screwdriver to tighten the screw securing the filament block to
the source mount.
5
Reinstall the ion source. (See
6
Close the front analyzer chamber. (See
7
Pumpdown the MS. (See
8
Autotune the MS. (