7
General Maintenance
To Install the EI XTR Source
7000/7010 Series TQ GC/MS Operating Manual
187
To Install the EI XTR Source
Materials needed
•
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
•
Tweezers (8710-2460)
•
Pliers, long-nose (8710-1094)
Procedure
1
Vent the MS. (See
2
Open the front analyzer chamber. (See
3
Slide the ion source into the source radiator. (See
4
Install and hand-tighten the source thumbscrews. Do not overtighten the
thumbscrews. (See
5
Connect the ion source wires. (See
“To Connect or Disconnect Wiring to the
CAUTION
Always wear clean gloves when working in the analyzer chamber to avoid
contamination.