CI Troubleshooting
3
Troubleshooting and Maintenance Manual
57
Figure 3
Schematic of CI flow control module
Gas A
(methane)
supply
Gas B
(other gas)
supply
Gas A
select valve
Gas B
select valve
Mass
flow
controller
Calibration
valve
Restrictor
Calibration
vial
Shutoff valve
GC/MS
interface
GC column
CI ion
source
Table 4
Flow module valve state diagram
Result
Gas A flow
Gas B flow
Purge
with Gas A
Purge
with Gas B
Pump out
flow module
Standby,
vented, or
EI mode
Gas A
Open
Closed
Open
Closed
Closed
Closed
Gas B
Closed
Open
Closed
Open
Closed
Closed
MFC
On (at
setpoint)
On (at
setpoint)
On (at 100%)
On (at 100%)
On (at 100%)
Off (at 0%)
Shutoff valve
Open
Open
Open
Open
Open
Closed