High mass sensitivity is poor
This refers to a condition where the
absolute
abundance at the upper end of
the mass range is poor. Absolute abundance should not be confused with the
relative
abundance (percentage) of
m/z
502 to
m/z
69. Sensitivity at high
masses can be excellent even if the relative abundance of
m/z
502 is low.
•
Wrong tune program
•
Wrong tune file
•
Repeller voltage is too low
•
Not enough time for the MS to warm up and pump down
•
Analyzer chamber pressure is too high
•
Column (carrier gas) flow is too high
•
Poor filament operation
•
Dirty ion source
•
Air leak
•
Incorrect DC polarity on the quadrupole mass filter
•
Collision cell gas flow is too high
•
Collision cell voltages are too high
•
Prefilter or postfilter
•
No voltage to the extractor lens (if using extractor EI source)