
GC Not Ready Symptoms
4
Troubleshooting
67
NPD Adjust Offset Process Fails
•
Inspect the jet to see if it is clogged.
•
Measure the actual detector flows. (See To Measure a
Detector Flow.)
•
Check the condition of the bead. Replace if necessary.
•
Verify that the flow settings are correct.
•
Thoroughly leak check the whole system, especially the
detector column fitting. (See Checking for Leaks.)
•
Set the equilibration time to 0.0.
Summary of Contents for 6890N
Page 1: ...Agilent Technologies Agilent 6890N Gas Chromatograph Troubleshooting ...
Page 20: ...20 Troubleshooting 2 Hardware Symptoms ...
Page 76: ...76 Troubleshooting 6 Nonfunctional GC Symptoms ...
Page 88: ...88 Troubleshooting 8 Troubleshooting Tasks ...
Page 102: ...102 Troubleshooting 8 Troubleshooting Tasks ...