GC Not Ready Symptoms
GC Never Becomes Ready
Normally the GC becomes ready after flows and temperatures
reach setpoint. If the GC does not become ready after a long
period of time:
•
Press
[Status]
or a component key (for example,
[Front inlet]
)
to see which setpoints or conditions are not ready.
•
Check for a sampler problem.
•
Check for a data system problem.
•
If performing manual injections in splitless or gas-saver
mode, you may need to press [
Prep Run
] to prepare the inlet
for the injection. Do this, for example:
•
To toggle the inlet purge valve before a splitless injection
•
To prepare for a pulsed injection
•
To turn off gas saver.