3 of 22
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Faults
850
Fault 205—Signal DSP RAM broken
Fault 205—Signal DSP RAM broken
Status message: Sig DSP RAM broken
The detector’s signal processing electronics are not functioning correctly.
This indicates a malfunction with the 6890 GC main PC board. The signal
path will not function.
❐
Turn the instrument on and then off at least one time. If the error still
occurs, the mainboard must be replaced.
Fault 206—Signal DSP registers
Status message: Sig DSP registers
The detector’s signal processing electronics are not functioning correctly.
This indicates a malfunction with the 6890 GC main PC board. The signal
path will not function.
❐
Turn the instrument on and then off at least one time. If the error still
occurs, the mainboard must be replaced.
Fault 207—Sig DSP data corrupt
Status message: Sig DSP data corrupt
The detector’s signal processing electronics are not functioning correctly.
This indicates a malfunction with the 6890 GC main PC board.
❐
Turn the instrument on and then off at least one time. If the error still
occurs, the mainboard must be replaced.
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......