830
Warning Codes
Warning 109—Front inlet: config changed, method defaulted
4 of 10
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Warning 109—Front inlet: config changed, method defaulted
Warning 110—Back inlet: config changed, method defaulted
Status message: F inl config changed or B inl config changed
This message appears when:
•
The 6890 GC is turned on after a new inlet has been installed.
•
The 6890 GC is turned on and senses that the inlet has a different type
of sensor than was there when the instrument was turned off.
•
A method is loaded that calls for a different type of inlet than the one
currently installed.
The following will occur:
❐
If a new inlet or sensor was installed, the 6890 GC will enter the default
setpoints for the new type of inlet.
❐
If a method was loaded with a different inlet specified, the 6890 GC will
not change the method but will flash the warning message.
Warning 111—Column 1: config changed, method defaulted
Warning 112—Column 2: config changed, method defaulted
Status message: Col 1 config changed or Col 2 config changed
This message appears when:
•
The 6890 GC is turned on after a different column has been installed.
•
You load a method that calls for a different type of column than the one
currently installed in the oven.
The following will occur:
❐
If a new column was installed, the 6890 GC will enter the column default
setpoints.
❐
If a method was loaded for a different type of column than the one
installed, the 6890 GC will not change the method but will flash the
warning message.
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......