520
Gases
The gas plumbing
8 of 14
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Traps
Using chromatographic-grade gases insures that the gas in your system is
pure. However, for optimum sensitivity, it is highly recommended that you
install high-quality traps to remove traces of water or other contaminants.
After installing a trap, check the gas supply lines for leaks.
Table 520-4
Recommended Traps
Moisture in carrier gas damages columns. We recommend a type 5A Molecular
Sieve trap after the source regulator and before any other traps.
A hydrocarbon trap removes organics from gases. It should be placed after a
molecular sieve trap and before an oxygen trap, if they are present.
An oxygen trap removes 99% of the oxygen from a gas plus traces of water. It
should be last in a series of traps. Because trace amounts of oxygen can
damage columns and degrade ECD performance, use an oxygen trap with
carrier and ECD gases. Do not use it with FID, FPD, or NPD fuel gases.
Figure 520-3
Traps
Description
Part No.
Preconditioned moisture trap: metal casing, s-shaped trap for carrier gas cleanup.
Contains Molecular Sieve 5A, 45/60 mesh, and 1/8-inch fittings.
5060-9084
Hydrocarbon trap: metal casing, s-shaped trap filled with 40/60 mesh activated
charcoal, and 1/8-inch fittings
5060-9096
Oxygen trap (for carrier and ECD gases): metal casing, and 1/8-inch brass fittings.
Oxygen trap cannot be reconditioned.
3150-0414
Molecular sieve or hydrocarbon trap—both are S-shaped
Oxygen trap
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......