Agilent 6890 Gas Chromatograph Service Manual
330
Thermal Conductivity Detector
(TCD)
Theory of operation
The TCD responds to any compounds whose thermal conductivity is different
from the thermal conductivity of the carrier gas alone. The TCD cell is a dual
channel device, with an empty flow path and a path containing a detector
filament. A switching valve alternates between sending the column effluent
(containing analytes) through the empty and the active flow paths. When the
column effluent flows through the empty channel, a pure stream of reference
gas maintains an equilibrium through the filament path. The reference gas is
used to compare thermal conductivity changes caused by the column effluent.
A gas with high thermal conductivity, such as helium, is used as the carrier/
make-up/reference gas. When the analyte is present in the gas stream, the
thermal conductivity drops, and less heat is lost to the cavity wall. Under
constant applied voltage, a silicon nitride coated filament in the TCD cell will
heat up and its resistance will increase. This change is what is recorded and
measured for the chromatogram.
Figure 330-1
The Thermal Conductivity Detector (TCD)
Pneumatic lines
Detector body
Heater/sensor cable
Filament leads
∆
PRT leads
(under cover)
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......