Nitrogen Phosphorus Detector (NPD)
320
Recommended settings
•
Turn the hydrogen flow off during solvent peaks and between runs.
•
If your NPD is
Off
for a long time in a high-humidity environment, water
may accumulate in your detector. To remove it:
a.
Set the detector temperature at 100°C and maintain it for 30 minutes.
b. Set the detector temperature to 150°C and maintain it for another
30 minutes.
Temperature programming
The NPD is flow sensitive. If you are using temperature programming, in
which the column flow resistance changes with temperature, set up the
instrument as follows:
•
With an EPC inlet, set the carrier gas in the
Constant flow
mode. Set
EPC detector makeup gas to
Const makeup.
NonEPC detectors provide
constant makeup gas flow.
•
If you have an EPC inlet and an EPC detector and choose to work in the
constant pressure mode, the makeup gas should be set in the
Col+mkup
=
const
mode.
Jets and collectors
The
capillary optimized
NPD is only used with capillary columns. It is shipped
with the standard jet and collector.
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......