
310
Flame Ionization Detector (FID)
Replacement procedures
4 of 24
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Replacement procedures
The detector is located underneath the plastic detector cover on the top of
the instrument. This section describes how to remove the entire detector as
well as how to remove individual parts.
Replacing the entire detector
WARNING
Before proceeding, turn off the oven and any heated zones and let them cool
down. Turn off any detector gases at their supply. Then turn off the main
power switch and unplug the power cord.
Caution
Make sure you are properly grounded with an ESD strap before continuing.
1.
Shut off the detector and the detector gases and let the detector cool.
2.
Remove the right side mainframe panel by loosening the two Torx T-20
captive screws.
3.
Disconnect the column from the bottom of the detector.
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......