Agilent 6890 Gas Chromatograph Service Manual
Solvent Vapor Exit Accessory
260
Leak testing
Leak testing
When checking the solvent vapor exit assembly for leaks, first check the
tri-column assembly. If the tri-column assembly is not leaking, check the
valve/fitting assembly (which includes the stainless steel union, tubing, and
fitting on the solenoid).
Potential leak points
•
Tri-column assembly
•
Quartz column connector
•
Quartz Y splitter
•
Connection to the inlet
•
Valve/fitting assembly
•
Solenoid valve
•
S/S union
•
Inlet fitting on the valve/fitting assembly
Note that there are no replacement parts on the valve/fitting assembly. If the
valve fails or the fitting cannot be tightened, replace the assembly.
Leak testing the tri-column assembly
Tools Required
The following tools (not supplied in the SVE Kit) are required to perform the
leak check procedure of the tri-column assembly.
•
1/16-inch (SS) union (part no. 0100-0124)
•
Vespel/graphite ferrules:
•
For 250
µ
m column (part no. 5062-3508)
•
For 320
µ
m column (part no. 5062-3506)
•
1/8-inch Swagelok nut (part no. 5180-4103)
•
1/8-inch solid (no hole) vespel ferrule (part no. 0100-1372) or
1/8-inch Swagelok cap
•
Electronic leak detector (optional)
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......