Agilent 6890 Gas Chromatograph Service Manual
Volatiles Interface
250
Theory of operation
Figure 250-4
VI flow control diagram: direct mode (idle or after sampling end)
Gas saver
Gas saver mode may be used in split or splitless operation mode. Gas saver
reduces the carrier gas flow from the split vent after the sample is on the
column. Column head pressure and the column flow rate are maintained,
while split vent flow decreases. Flows remain at the reduced level until the
beginning of the next run.
Sampler
Trap
Valve
open
PV2
SPR
SPR
Vent
Split
Vent
FS
Supply
To Column
PS