Agilent 6890 Gas Chromatograph Service Manual
Programmed Temperature Vaporization Inlet (PTV)
240
Replacement procedures
8.
Check all connections for leaks. If necessary, the retaining nut can be
tightened an additional 1/4 turn to eliminate leaks.
Replacing the septum
Either a regular septum or a Merlin microseal can be used with the septum
head.
If the inlet temperature is set below 40°C, the Merlin microseal may not seal
effectively. For inlet temperatures below 40°C, use a regular septum for the
inlet seal.
1.
To replace the septum, cool the inlet to ambient temperature.
2.
Unscrew (counterclockwise) the septum cap or Merlin cap.
3.
Remove the septum or Merlin microseal, taking care not to scratch the
interior of the septum head.
4.
Install a new septum or Merlin microseal and the correct cap. When
installing a Merlin microseal, note that the side with the visible metal
parts faces the oven.
Figure 240-23
PTV septa
5.
Check for leaks out of the cap and tighten the cap if necessary.
Replacing the liner
1.
Remove the head from the inlet.
2.
Grasp the liner by the Graphpack ferrule. Remove the liner and ferrule.
Summary of Contents for 6890 Series
Page 1: ...6890 Series Gas Chromatograph Service Manual ...
Page 2: ...6890 Series Gas Chromatograph Service Manual ...
Page 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Page 29: ......
Page 39: ......
Page 207: ......
Page 391: ......
Page 459: ......
Page 493: ......
Page 521: ......
Page 523: ......
Page 580: ...900 Remote Access Remote Access information was not available for this printing ...
Page 581: ......
Page 583: ......
Page 631: ......
Page 677: ......
Page 737: ......