Compatibility Language - B
85
Table B-2. COMPatibility Commands (continued)
Compatibility
Command
Description
Similar SCPI
Command
OVSET <n>
This command programs the overvoltage protection. See Table
4-3 for the programming range of this command.
Initial condition: MAX
VOLT:PROT
POL 0 | 1
Only applies to units with Option 760. This command sets the
polarity of the output relays to either normal (1) or inverted (0).
Initial condition: POL 1
OUTP:REL:POL 0|1
PON 0 | 1
This command enables (1) or disables (0) SRQ at power-on.
Initial condition: last programmed value
PSC 0 | 1
RELAY 0 | 1
Only applies to units with Relay Option 760. This command opens
(0) or closes (1) the output relays without affecting the
programmed output state of the unit.
Initial condition: RELAY 1
OUTP:REL 0 | 1
RLYPON 0 | 1
Only applies to units with Relay Option 760. This command opens
(0) or closes (1) the output relays at power-on without affecting
the programmed output state of the unit.
Initial condition: RLYPON 1
RCL 0
ROM?
This command returns the ROM version of the dc source.
Data Representation: AAA AAA
*IDN?
RST
This command resets the dc source if the output is disabled by
the output protection circuits.
OUTP:PROT:CLE
SENS:CURR
:RANG <n>
This command sets the current measurement range of the dc
source. See Table 4-3 for the programming range of this
command.
Initial condition: MAX
SENS:CURR:RANG
SENS:SWE
:POIN <n>
This command defines the number of data points in a
measurement sweep.
Initial condition: 32
SENS:SWE:POIN
SENS:SWE
:TINT <n>
This command defines the time period between measurement
samples.
Initial condition: 15.6 µs.
SENS:SWE:TINT
SRQ 0 | 1
These commands enable or disable the dc source's ability to
request service from the controller for fault conditions. UNMASK
defines which conditions are defined as faults.
Initial condition: SRQ 0
*SRE
STS?
This command reads the contents of the status register, which
maintains the present status of the dc source.
Data Representation: ZZZZD
STAT:OPER:COND?
STAT:QUES:COND?
*ESE?
SYST:LANG
This command causes the alternate language to become active
and to be stored in nonvolatile memory. In this case, the
commands are equivalent. After being shut off, the dc source will
resume in the last-selected language when power is restored.
The parameter must be either COMP or SCPI.
SYST:LANG
TEST?
This command causes the dc source to run selftest and report
any detected failures.
Data Representation: ZZZZD
*TST?