5977B Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
Maintaining the Vacuum System
Periodic maintenance
Some maintenance tasks for the vacuum system must be performed
periodically. (See
on page 91.) These include:
•
Checking the foreline pump fluid (every week)
•
Checking the calibration vial (every 6 months)
•
Replacing the foreline pump oil (every 6 months; every 3 months for
CI MSDs using ammonia reagent gas, standard foreline pump)
•
Tightening the foreline pump oil box screws (first oil change after
installation, standard foreline pump)
•
Changing the foreline pump exhaust filters
•
Replacing the DP fluid (once a year)
•
Changing the dry foreline pump tip seal (once a year)
Failure to perform these tasks as scheduled can result in decreased
instrument performance. It can also result in damage to your instrument.
Other procedures
Tasks such as replacing a Micro-Ion vacuum gauge should be performed only
when needed. (See
on page 35.) Refer to the online
help in the Agilent MassHunter GC/MS Acquisition software for symptoms
that indicate this type of maintenance is required.
More information is available
If you need more information about the locations or functions of vacuum
system components. (See
Most of the procedures in this chapter are illustrated with video clips on the
5977B Series MSD User Information media.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......