![Agilent Technologies 5977B Series Troubleshooting And Maintenance Manual Download Page 65](http://html.mh-extra.com/html/agilent-technologies/5977b-series/5977b-series_troubleshooting-and-maintenance-manual_2867608065.webp)
CI Troubleshooting
3
65
Troubleshooting Tips and Tricks
Rule 1: Look for what has been changed.
Many problems are introduced accidentally by human actions. Every time any
system is disturbed, there is a chance of introducing a new problem.
•
If the MSD was just pumped down after maintenance, suspect air leaks or
incorrect assembly.
•
If the reagent gas bottle or gas purifier were just changed, suspect leaks or
contaminated or incorrect gas.
•
If the GC column was just replaced, suspect air leaks or contaminated or
bleeding column.
•
If you have just switched ion polarity or reagent gas, suspect the tune file
you have loaded in memory. Is it the appropriate file for your mode of
operation?
Rule 2: If complex is not working, go back to simple.
A complex task is not only more difficult to perform, but also more difficult to
troubleshoot as well. For example, CI requires more parts to work correctly
than EI does.
•
If you are having trouble with NCI, verify that PCI still works.
•
If you are having trouble with other reagent gases, verify that methane still
works.
•
If you are having trouble with CI, verify that EI still works.
Rule 3: Divide and conquer.
This technique is known as half-split troubleshooting. If you can isolate the
problem to only part of the system, it is much easier to locate.
•
To isolate an air leak, select
Shutoff valve
. If the abundance of
m/z
32
decreases, the problem is not in the flow module.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......