General Troubleshooting
2
5977B Series MSD Troubleshooting and Maintenance Manual
61
Contamination
Contamination is usually identified by excessive background in the mass
spectra. It can come from the GC or from the MSD. The source of the
contamination can sometimes be determined by identifying the contaminants.
Some contaminants are much more likely to originate in the GC. Others are
more likely to originate in the MSD.
Contamination originating in the GC typically comes from one of these
sources:
•
Column or septum bleed
•
Dirty GC inlet
•
GC inlet liner
•
Contaminated syringe
•
Poor quality carrier gas
•
Dirty carrier gas tubing
•
Fingerprints (improper handling of clean parts)
Contamination originating in the MSD typically comes from one of the
following sources:
•
Air leak
•
Cleaning solvents and materials
•
Foreline pump oil (standard foreline pump)
•
Fingerprints (improper handling of clean parts)
on page 62 lists some of the more common contaminants, the ions
characteristic of those contaminants, and the likely sources of those
contaminants.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......