General Troubleshooting
Air Leaks
Air leaks are a problem for any instrument that requires a vacuum to operate.
Leaks are generally caused by vacuum seals that are damaged or not fastened
correctly. Symptoms of leaks include:
•
Higher than normal analyzer chamber pressure or foreline pressure
•
Higher than normal background
•
Peaks characteristic of air (
m/z
18, 28, 32, and 44 or
m/z
14 and 16)
•
Poor sensitivity
•
Low relative abundance of
m/z
502 (this varies with the tune program used)
Leaks can occur in either the GC or the MSD. The most likely point for an air
leak is a seal you recently opened.
In the GC, most leaks occur in:
•
GC inlet septum
•
GC inlet column nut
•
Broken or cracked capillary column
Leaks can occur in many more places in the MSD:
•
GC/MSD interface column nut
•
Side plate O-ring (all the way around)
•
Vent valve O-ring
•
Calibration valve(s)
•
GC/MSD interface O-ring (where the interface attaches to the analyzer
chamber)
•
Front and rear end plate O-rings
•
Turbo pump O-ring
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......