DC polarity
The DC polarity (DC Pol) parameter selects the orientation of the direct
current applied to the quadrupole mass filter. The DC Pol that works best for
your MSD is determined at the factory. It is listed on the final test sheet
accompanying your MSD. It is also listed on a label on the cover over the RF
coils. This cover can be viewed by removing the upper MSD cover.
Mass (axis) gain
Mass gain (MassGain) controls the mass assignment, that is, assignment of a
particular peak to the correct
m/z
value.
•
A higher gain yields higher mass assignment.
•
Mass gain affects peaks at high masses more than peaks at low masses.
Mass (axis) offset
Mass offset (MassOffs) also controls the mass assignment.
•
A higher offset yields higher mass assignment.
•
Mass offset generally affects peaks equally at all masses.
Quadrupole maintenance
The mass filter requires no periodic maintenance. It should not be removed
from the radiator. If
absolutely
necessary (that is, if the only alternative is
replacement), the quadrupole can be cleaned. Cleaning
must
be performed by
Agilent Technologies service personnel.
C A U T I O N
Using the nonpreferred DC polarity may result in very poor performance. Always use
the factory-specified polarity.
C A U T I O N
Never put the quadrupole in an ultrasonic cleaner.
Never change the physical orientation of the quadrupole mass filter.
The fused-quartz quadrupole is fragile and will break if dropped or handled roughly.
The material in the cusps of the quadrupole is very hygroscopic. If exposed to water,
the quadrupole must be dried very slowly to prevent damage.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......