![Agilent Technologies 5977B Series Troubleshooting And Maintenance Manual Download Page 137](http://html.mh-extra.com/html/agilent-technologies/5977b-series/5977b-series_troubleshooting-and-maintenance-manual_2867608137.webp)
General Maintenance
4
5977B Series MSD Troubleshooting and Maintenance Manual
137
To Lubricate the Side Plate O-Ring
Materials needed
•
Cloths, clean (05980-60051)
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Grease, Apiezon L, high vacuum (6040-0289)
The side plate O-ring needs a thin coat of grease to ensure a good vacuum seal.
If the side plate O-ring appears dry or does not seal correctly, lubricate it using
this procedure. A good test is to wipe off the side plate with methanol, then
close the analyzer chamber. If the O-ring has enough grease on it, it will leave a
faint trace on the side plate.
Procedure
1
Vent the MSD.
2
Open the analyzer chamber.
3
Use a clean, lint-free cloth or glove to spread a
thin
coat of high vacuum
grease only on the exposed surface of the O-ring. (See
4
Use a clean, lint-free cloth or glove to wipe away excess grease. If the
O-ring looks shiny, there is too much grease on it.
5
Close the analyzer chamber.
6
Pump down the MSD.
C A U T I O N
Vacuum seals other than the side plate O-ring and vent valve O-ring do not need to be
lubricated. Lubricating other seals can interfere with their correct function.
C A U T I O N
Do not use anything except the recommended vacuum grease. Excess grease can trap
air and dirt. Grease on the surface of the O-ring other than the exposed surface can trap
air, resulting in air spikes during operation.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......