To Replace the Fan for the High Vacuum Pump
Materials needed
Fan (G7005-60564)
Screwdriver, Torx T-20 (8710-1615)
Procedure
1
Vent the MSD.
2
Remove the left side MSD cover.
3
Disconnect the fan wiring from the connector on the MSD frame. (See
4
Remove the four fan screws and the safety grill. Remove the fan. Keep the
screws.
5
Install the new fan with the flow arrow on the side pointing toward the
pump. The wires should be at the upper left, close to the connector.
6
Add the safety grill and the four screws. Tighten the screws firmly.
7
Connect the fan wiring to the fan connector on the MSD frame.
8
Reinstall the MSD covers.
9
Pump down the MSD.
WA R N I N G
Do not touch the high vacuum pump. The pump could still be hot enough to burn you.
WA R N I N G
Ensure the safety grill that shields the fan blades is in place.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......