Installing Intuvo 9000 GC Columns
To Install the GC/MSD Interface Tip Seal
5977B Series MSD Operating Manual
75
3
Thread the knurled tip seal retainer into the tip seal base finger tight.
4
Gently
check the alignment of the analyzer and the interface.
When the analyzer is aligned correctly, the analyzer can be closed all the
way with no resistance except the spring tension from the interface tip seal.
5
You can align the analyzer and interface by wiggling the side plate on its
hinge. If the analyzer still will not close, contact your Agilent Technologies
service representative.
Figure 17.
GC/MSD interface tip seal
Interface tip seal
Knurled tip seal retainer
Tip seal base
CAUTION
Forcing the analyzer closed if these parts are misaligned will damage the seal
or the interface or the ion source, or keep the sideplate from sealing.