7
General Maintenance
To Assemble the EI XTR Source
212
5977B Series MSD Operating Manual
To Assemble the EI XTR Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Hex ball driver, 2.0 mm (8710-1804)
•
Wrench, open-end, 10 mm (8710-2353)
Procedure
Refer to the exploded parts view
and EI XTR parts list
page 213 while using this procedure.
1
Slide the ceramic washer into the source body.
2
Insert the extractor lens into the source body, flat side first.
3
Insert the entrance lens and ion focus lens into the insulator in the order
shown in
4
Slide the insulator containing the ion focus and entrance lens into the source
body, with the ion focus lens against the extractor lens.
5
Install the two gold plated setscrews that holds the lenses in place.
6
Assemble the repeller assembly.
a
Install the repeller block insert into the source heater block assembly.
b
Install the repeller insulators into the source heater block assembly and
repeller block insert.
c
Install the repeller through the repeller insulators, then put the flat
washer followed by the belleville spring washer onto the end of the
repeller shaft and secure finger tight with the repeller nut.
7
Attach the repeller assembly to the source body using the two gold plated
screws and spring washers.
8
Install the filaments using the two gold plated screws and spring washers.
CAUTION
Do not overtighten the repeller nut or the ceramic repeller insulators will break
when the source heats up. The nut should only be finger tight.
Summary of Contents for 5977B CI MSD
Page 1: ...Agilent 5977B Series MSD Operating Manual ...
Page 12: ...12 5977B Series MSD Operating Manual ...
Page 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Page 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Page 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Page 253: ......