Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
T6 Torx driver, (8710-2548)
•
Tweezers (8710-2460)
Procedure
Refer to the exploded parts view
page 190 while using this procedure.
1
Assemble the five lenses inside the lens insulator. (See
page 188.) The lens number is engraved into the outer circumference of each
lens.
a
Starting with the entrance lens 1, set the lens into the end groove in the
lens insulator, and rotate the lens until you feel the ball seat into the
circular recess.
b
Insert the next 4 lenses, in numerical order, into the lens insulator. The
lens chamber open end always faces the entrance lens 1. Rotate each
lens until you feel the ball seat into the circular recess.
It is easier to insert lens 5 on an angle because the lens stack at this
point makes the lens insulator less flexible.
CAUTION
Always wear clean gloves when working in the analyzer chamber to avoid
contamination.
CAUTION
Use care when inserting the lenses into the lens insulator casing. Putting
excessive stress on this casing can break or crack it. If this happens, do not
attempt to operate with a defective lens insulator, it must be replaced.
Summary of Contents for 5977B CI MSD
Page 1: ...Agilent 5977B Series MSD Operating Manual ...
Page 12: ...12 5977B Series MSD Operating Manual ...
Page 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Page 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Page 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Page 253: ......