5
Operating in EI Mode
Operating the MSD from the 7890 GC Control Panel
106
5977B Series MSD Operating Manual
To view the firmware version of the MSD from the GC control panel
1
Press
MS/Aux Det
to bring up the 5977B MSD menu.
2
Press the down arrow to scroll to
Firmware
.
To view the serial number of the MSD from the GC control panel
1
Press
MS/Aux Det
to bring up the 5977B MSD menu.
2
Press the down arrow to scroll to
Serial#
.
To configure the network settings for the MSD from the GC control panel
1
Press
Config,
and then press
MS/Aux Det
to bring up the CONFIGURE MS
DETECTOR menu.
2
To configure the
IP:
parameter, use the GC keypad to enter the new IP
address for the MSD, then press
Enter
to complete the entry.
3
Wait for the GC to display the new IP address. Reboot the MSD or proceed to
the gateway address with the down arrow.
4
Press the down arrow to scroll to
GW:
and use the GC keypad to enter the
new gateway address for the LAN and press
Enter
to complete the entry.
5
Press the down arrow to scroll to
SW:
and use the GC keypad to enter the
new subnet mask for the LAN and press
Enter
to complete the entry.
6
Reboot the MSD. (See below.)
To reboot the MSD from the GC control panel
1
Press
Config,
and then press
MS/Aux Det
to bring up the CONFIGURE MS
DETECTOR menu.
2
Press the down arrow to scroll to
Request MSD Reboot?
.
3
Press
On/Yes
to reboot the MSD and wait for the MSD to complete this cycle
before trying to access it. The LED turns red until the reboot cycle is
completed.
Summary of Contents for 5977B CI MSD
Page 1: ...Agilent 5977B Series MSD Operating Manual ...
Page 12: ...12 5977B Series MSD Operating Manual ...
Page 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Page 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Page 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Page 253: ......