![Agilent Technologies 5975T LTM GC/MSD Troubleshooting And Maintenance Manual Download Page 88](http://html1.mh-extra.com/html/agilent-technologies/5975t-ltm-gc-msd/5975t-ltm-gc-msd_troubleshooting-and-maintenance-manual_2867607088.webp)
88
Troubleshooting and Maintenance Guide
6
Shutdown Symptoms
Column Shutdowns
The instrument monitors inlet and auxiliary gas streams. If a
carrier gas (which can include an auxiliary flow module or
pneumatics control module) is unable to reach its flow or
pressure setpoint, the instrument assumes that a leak exists. It
will warn you with a beep after 25 seconds, and it will continue
to beep in intervals. After about 5 minutes, the instrument will
shut down components to create a safe state. The instrument:
•
Displays
Front inlet pressure shutdown
.
•
Turns off to avoid column damage.
•
Flashes LTM column temperature setpoint
Off
.
•
Turns off all flows for the column. When viewed, their
parameters flash
Off
. For example, the septum purge and
column flows would turn off.
•
Turns off all other heaters. When viewed, their temperature
parameters flash
Off
.
•
Attempts to turn on a shut-down zone fail with an error
message.
To recover from this state.
1
Fix the cause of the shutdown.
•
Check for a broken column near the inlet.
•
Check for leaks.
•
Replace the inlet septum.
•
Replace the inlet O-ring.
•
Check the supply pressure.
2
Press the key for the device that initiated the shutdown.
Scroll to the pneumatic parameter that is flashing
Off
, then
press
[On]
or
[Off]
.
For example, if the inlet ran out of carrier gas, click
[Inlet]
from the electronic keypad, scroll to the pressure or flow
parameter, then click
[On]
.
Summary of Contents for 5975T LTM GC/MSD
Page 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Page 10: ...10 Troubleshooting and Maintenance Guide...
Page 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Page 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Page 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Page 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Page 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Page 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Page 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...