ALS Symptoms
5
Troubleshooting and Maintenance Guide
85
Syringe Needle Bends During Injection into Inlet
Refer to your
7693 Automatic Liquid Sampler Installation,
Operation and Maintenance Manual
for additional
information.
•
Check that the GC septum nut is not too tight.
•
Check that the syringe is installed correctly into the syringe
carriage.
•
Check that the needle support and guide are clean. Remove
any residue or septum deposits.
•
Check that you are using the proper syringe. The combined
length of the syringe barrel and needle should be
approximately 126.5 mm.
WA R N I N G
When troubleshooting the injector, keep your hands away from
the syringe needle. The needle is sharp and may contain
hazardous chemicals.
Summary of Contents for 5975T LTM GC/MSD
Page 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Page 10: ...10 Troubleshooting and Maintenance Guide...
Page 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Page 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Page 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Page 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Page 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Page 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Page 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...