Troubleshooting and Maintenance Guide
3
Chromatographic Symptoms
Low Peak Area or Height (Low Sensitivity)
•
If using an inlet in split mode, check the split ratio.
•
Check for leaks. (See
•
Check the inlet for contamination. (See
•
Check each column and verify that it was cut and installed
properly at each end.
•
Verify that the column type is correct.
•
Perform column maintenance: Bake out contaminants,
replace the inlet guard column, and reverse and bake out the
column as needed.
•
Verify that the liner type is appropriate for the sample.
•
Check the supply gas purity.
•
Check all trap indicators and dates.
•
Verify that the method parameters are correct.
•
Check sample stability.
•
Check configured syringe size. Some syringe sizes are
specified at half-capacity. If the maximum syringe volume is
marked at half-height on the barrel, not at the top of the
barrel, enter
twice
the labeled volume when configuring the
syringe size.
Summary of Contents for 5975T LTM GC/MSD
Page 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Page 10: ...10 Troubleshooting and Maintenance Guide...
Page 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Page 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Page 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Page 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Page 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Page 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Page 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...