Chromatographic Symptoms
3
Troubleshooting and Maintenance Guide
35
Retention Times Not Repeatable
•
Replace the septum. (See the
5975T LTM GC/MSD Operation
Manual.
)
•
Check for leaks in the inlet, liner (as applicable), and
column connection. (See
•
Check for sufficient carrier gas supply pressure. The
pressure delivered to the instrument must be at least 40 kPa
(10 psi) greater than the maximum inlet pressure required
at final LTM column temperature.
•
Run replicates of known standards to verify the problem.
•
Verify that you are using the correct liner type for the
sample being injected.
•
Consider if this is the first run. (Has the instrument
stabilized?)