Analyzer
7
5975 Series MSD Troubleshooting and Maintenance Manual
209
AMU gain
AMU gain (AmuGain) affects the ratio of DC voltage to RF frequency on the
mass filter. This controls the widths of the mass peaks.
•
Higher gain yields narrower peaks.
•
AMU gain affects peaks at high masses more than peaks at low masses.
AMU offset
AMU offset (AmuOffs) also affects the ratio of DC voltage to RF frequency on
the mass filter.
•
Higher offset yields narrower peaks.
•
AMU offset generally affects peak widths equally at all masses.
219 width
m/z
219 is a prominent ion near the middle of the mass range of PFTBA. The
width parameter (Wid219) makes small corrections to the
m/z
219 peak width.
Amu gain and amu offset must be readjusted after the 219 width is changed. If
you are tuning with a compound other than PFTBA, there may not be an ion at
m/z
219. In that case, set the 219 width to the last value found for it by
Autotune or set it to 0.
Figure 47
Quadrupole mass filter
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...