Filament care
Like the filaments in incandescent light bulbs, the ion source filaments will
eventually burn out. Certain practices will reduce the chance of early failure:
•
If you have an optional G3397A Micro-Ion vacuum gauge, use it to verify
that the system has an adequate vacuum before turning on the analyzer,
especially after any maintenance was performed.
•
If you are controlling your MSD from the Edit Parameters screen, always
select MSOff before changing any of the filament parameters.
•
When setting up data acquisition parameters, set the solvent delay so that
the analyzer will
not
turn on while the solvent peak is eluting.
•
When the software prompts Override solvent delay? at the beginning of a
run,
always
select NO.
•
Higher emission current will reduce filament life.
•
Higher electron energy will reduce filament life.
•
Leaving the filament on for short times (
1
minute) during data acquisition
will reduce filament life.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...