Introduction
1
5975 Series MSD Troubleshooting and Maintenance Manual
19
Many parts are dangerously hot
Many parts of the GC/MSD operate at temperatures high enough to cause
serious burns. These parts include but are not limited to:
•
The inlets
•
The oven and its contents
•
The detector
•
The column nuts attaching the column to an inlet or detector
•
The valve box
•
The foreline pump
Always cool these areas of the system to room temperature before working on
them. They will cool faster if you first set the temperature of the heated zone
to room temperature. Turn the zone off after it has reached the setpoint. If you
must perform maintenance on hot parts, use a wrench and wear gloves.
Whenever possible, cool the part of the instrument that you will be
maintaining before you begin working on it.
The oil pan under the standard foreline pump can be a fire hazard
Oily rags, paper towels, and similar absorbents in the oil pan could ignite and
damage the pump and other parts of the MSD.
WA R N I N G
Be careful when working behind the instrument. During cool-down cycles, the GC
emits hot exhaust which can cause burns.
WA R N I N G
The insulation around the inlets, detectors, valve box, and the insulation cups is
made of refractory ceramic fibers. To avoid inhaling fiber particles, we recommend
the following safety procedures: ventilate your work area; wear long sleeves,
gloves, safety glasses, and a disposable dust/mist respirator; dispose of insulation
in a sealed plastic bag; wash your hands with mild soap and cold water after
handling the insulation.
WA R N I N G
Combustible materials (or flammable/non-flammable wicking material) placed
under, over, or around the foreline (roughing) pump constitutes a fire hazard. Keep
the pan clean, but do not leave absorbent material such as paper towels in it.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...