CI Maintenance
5
5975 Series MSD Troubleshooting and Maintenance Manual
169
To Refill the CI Calibrant Vial
Materials needed
•
PFDTD calibrant (8500-8510)
Procedure
1
Set the reagent gas flow to
Gas Off.
2
Vent the MSD.
3
Remove the capillary column from the GC/MSD interface.
4
Pull the MSD away from the GC to expose the calibration vial and valve.
See
5
Loosen the collar holding the calibration vial in place. Do not remove the
collar.
6
Remove the calibrant vial. See
7
Fill the vial no higher than the bottom of the internal tube with fresh
PFDTD calibrant (8500-8510).
8
Replace the vial and tighten the collar.
9
Reposition the MSD next to the GC. See
.
10
Reinstall the capillary column.
11
Pump down the MSD.
12
Purge the calibration valve. See
C A U T I O N
Do
not
rinse the vial with any solvents.
Never
expose the inside of the vial to
chlorinated solvents or isopropyl alcohol or water — this will result in severe loss of CI
sensitivity.
C A U T I O N
After removing the calibrant vial, you
must
purge the calibration valve. Failure to do so
will result in severe contamination of the ion source and damage to the filament and
electron multiplier.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...