168
5975 Series MSD Troubleshooting and Maintenance Manual
5
CI Maintenance
To Clean the Reagent Gas Supply Lines
Materials needed
•
Clean, dry nitrogen
•
Heat gun
•
Tubing cutter (8710-1709)
Procedure
If the reagent gas lines become contaminated, they can be cleaned.
1
Disconnect the reagent gas tubing from the gas supply, the gas purifier, and
the MSD.
2
Cap the gas purifier following the instructions on the label.
3
Connect one end of the tubing to a supply of clean, dry nitrogen and turn
on gas flow.
4
Use the heat gun to warm the tubing, starting at the supply end and
working your way to the free end.
5
Repeat for any other pieces of tubing that need to be cleaned.
6
Reconnect the tubing to the gas supply, gas purifier, and MSD. Follow the
instructions on the gas purifier label.
WA R N I N G
Do not heat the gas tubing when reagent gas is flowing.
C A U T I O N
Do not put liquids into the tubing. Do not heat the tubing when it is connected to the
MSD.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...