General Maintenance
4
5975 Series MSD Troubleshooting and Maintenance Manual
151
To replace the electron multiplier horn
Materials needed
•
Electron multiplier horn (G3170-80103)
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
Procedure
1
Vent the MSD.
2
Open the analyzer chamber.
3
Open the retaining clip (
). Lift the arm of the clip up and then
swing the clip away from the electron multiplier horn.
4
Remove the electron multiplier horn.
5
Install the new electron multiplier horn.
6
Close the retaining clip.
The signal pin on the horn must rest
on the outside
of the loop in the
contact strip.
Do not
put the signal pin on the inside of the loop in the
contact strip. Incorrect installation will result in poor sensitivity or no
signal.
7
Close the analyzer chamber.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...