General Maintenance
4
5975 Series MSD Troubleshooting and Maintenance Manual
133
To disassemble the EI ion source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Hex ball driver, 2.0 mm (8710-1804)
•
Wrench, open-end, 10 mm (8710-2353)
Procedure
1
Remove the ion source. See the
Operation Manual
.
2
Remove the filaments (
3
Separate the repeller assembly from the source body. The repeller
assembly includes the source heater assembly, repeller, and related parts.
4
Remove the repeller.
5
Unscrew the interface socket. A 10-mm open-end wrench fits the flats on
the interface socket.
6
Remove the setscrew for the lenses.
7
Push the lenses out of the source body.
N O T E
Video shows the standard Ion Source Assembly procedures.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...