5975 Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
To Replace the Fan for the High Vacuum Pump
Materials needed
•
Fan (G1099-60564)
•
Screwdriver, Torx T-15 (8710-1622)
Procedure
1
Vent the MSD.
2
Remove the left side MSD cover.
3
Disconnect the fan wiring from the connector on the MSD frame
(
4
Remove the four fan screws and the safety grill. Remove the fan. Keep the
screws.
5
Disconnect the fan wiring from the old fan. The fan wiring ends in a small
connector on the back of the fan.
6
Connect the fan wiring to the new fan.
7
Install the new fan with the flow arrow on the side pointing toward the
pump. The wires should be at the upper left, close to the connector.
8
Add the safety grill and the four screws. Tighten the screws firmly.
9
Connect the fan wiring to the fan connector on the MSD frame.
10
Reinstall the MSD covers.
11
Pump down the MSD.
WA R N I N G
Do not touch the high vacuum pump. The pump could still be hot enough to burn you.
WA R N I N G
Make sure the safety grill that shields the fan blades is in place.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...