Installing 7890/7820/6890/6850 GC Columns
5975/5977 Series MSD Operating Manual
41
To Install a Capillary Column in the GC/MS
Interface Using a Standard Column Nut
This procedure is for the installation of a capillary column directly into the
analyzer. There are two types of column nuts that can be used in the GC/MS
interface: The standard column nut explained here, and the self tightening
column nut explained in
“To Install a Capillary Column in the GC/MS Interface
Using the Self-Tightening Column Nut”
Materials needed
•
Interface tip seal (G3870-20542) (See
page 38.) (Not used with EI SS or Inert source)
•
Tip seal spring (G7005-20024) (Not used with SS or Inert EI source)
•
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
•
Flashlight
•
Magnifying loupe
•
Gloves, clean
•
Large (8650-0030)
•
Small (8650-0029)
•
Interface column nut (05988-20066)
•
Ferrules
•
0.3 mm id, for 0.10 mm id columns (5062-3507)
•
0.4 mm id, for 0.20 and 0.25 mm id columns (5062-3508)
•
0.5 mm id, for 0.32 mm id columns (5062-3506)
•
0.8 mm id, for 0.53 mm id columns (5062-3512)
•
Septum (may be old, used inlet septum)
•
Safety glasses
•
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)