Chapter 8 Ensuring Repeatability, Accuracy, and Resolution
Ensuring Accuracy
Getting Started Guide
8-9
8
Correcting errors that affect angular measurements
The accuracy of your angular measurement can be reduced by even a
slight change in the distance between the retroreflectors. This distance
changes if the temperature of the angular reflector’s housing changes
during the measurement. To minimize this error, follow these guidelines:
•
Avoid excessive handling of the angular reflector.
•
Avoid moving the angular reflector over heated or cooled surfaces.
•
Avoid having heated or cooled air moving past the angular reflector.
Correcting errors that affect straightness, squareness,
and parallelism measurements
The accuracy of your straightness measurement can be reduced by the
following factors:
•
the accuracy of the straightness reflector mirrors
•
misalignment (slope) between the machine travel and the laser
measurement axis
•
environmental factors
Ensuring the accuracy of straightness reflector mirrors
The accuracy of the straightness measurement depends on the two plane
mirrors in the straightness reflector having the same flatness. If one
mirror is convex or concave, the system will misinterpret this as an out-of-
straightness condition.
To minimize the effect of the mirrors along the horizontal axis, follow
these steps:
1
Rotate the straightness reflector 180 degrees and make a second set of
measurements.
2
Average the data collected for each point.
See the Agilent 5530 Dynamic Calibrator Measurements Reference Guide
for more information.
Summary of Contents for 5530
Page 1: ...Agilent 5530 Dynamic Calibrator Getting Started Guide ...
Page 2: ......
Page 3: ...Agilent 5530 Dynamic Calibrator Getting Started Guide ...
Page 10: ...Contents viii Getting Started Guide ...
Page 13: ...1 Introducing the Agilent 5530 ...
Page 20: ...Chapter 1 Introducing the Agilent 5530 Types of Measurements 1 8 Getting Started Guide 1 ...
Page 21: ...2 Making Laser Measurements Safely ...
Page 26: ...Chapter 2 Making Laser Measurements Safely Protecting the Optics 2 6 Getting Started Guide 2 ...
Page 27: ...3 Installing the Software and Connecting Components ...
Page 45: ...4 Using the Agilent 5530 for the First Time ...
Page 66: ......
Page 67: ...5 Setting Up the Software for a Measurement ...
Page 91: ...6 Making the Measurement ...
Page 101: ...7 Analyzing Transferring and Printing Measurement Data ...
Page 113: ...8 Ensuring Repeatability Accuracy and Resolution ...
Page 127: ...9 Troubleshooting and Maintenance ...
Page 133: ...Glossary ...
Page 138: ...Glossary G 6 Getting Started Guide Glossary ...
Page 144: ...Index Index 6 Getting Started Guide ...
Page 149: ......
Page 150: ... 10747 90061 Manual Part Number 10747 90061 Printed in U S A OCTOBER 13 2008 ...