Using the Laser Head in a System
Solving Problems
Here is information about three simple problems that may occur while you
are trying to make a measurement.
•
If there is no red beam from the laser head:
•
Ensure that the laser head cable is securely connected to the back of
the laser head.
•
Ensure that the power supply is connected to the laser head cable.
•
Ensure that the power supply is plugged into an operating outlet and
that the
POWER ON
light is illuminated.
•
If the green LED on the receiver does not light:
•
Ensure that there is a red beam from the laser head and the
Ready
LED is illuminated.
•
Ensure that the laser beam passes through the large aperture in the
laser head.
•
Ensure the laser head’s reference connector is connected to the
electronics.
•
Ensure that the receiver cable is securely connected at the receiver
and at the system electronics.
•
Ensure that the system electronics are connected to an operating
power source and are turned on.
•
Ensure that the alignment of the laser head, optics, and receiver
causes the beam from the interferometer to be centered on the
receiver´s lens.
•
Block the beam between the interferometer and measurement
reflector. The reference- path beam should be centered on the
receiver’s lens. When you unblock the measurement- path beam, it
should overlap the reference- path beam on the receiver lens.
•
If everything above seems correct, but the LED still doesn’t light,
adjust the receiver’s gain control.
•
If the controller does not indicate motion of the measurement reflector:
•
Ensure that there is a red beam from the laser head and the
Ready
LED is illuminated.
•
Ensure that the laser system electronics are powered.
•
Ensure that the laser beam is properly aligned and that the
receiver’s green LED is lighted.
•
If the receiver’s green LED remains lighted when no measurement
signal is present, reduce the receiver’s gain via its gain control.
•
Ensure that the controller is on and the program is running.
Summary of Contents for 5517B
Page 1: ...Agilent Technologies Agilent 5517B BL C D DL FL Laser Head User s Guide...
Page 10: ...x Agilent 5517B BL C D DL FL User s Guide Contents...
Page 12: ...xii Agilent 5577B BL C D DL FL User s Guide Figures...
Page 14: ...xiv Agilent 5577B BL C D DL FL User s Guide Tables...
Page 24: ...10 Agilent 5517B BL C D DL FL User s Guide 1 General Information...
Page 28: ...14 Agilent 5517B BL C D DL FL User s Guide 2 Laser Head at a Glance...
Page 52: ...38 Agilent 5517B BL C D DL FL User s Guide 4 Performance Tests...
Page 64: ...50 Agilent 5517B BL C D DL FL User s Guide 5 Adjustments...
Page 86: ...72 Agilent 5517B BL C D DL FL User s Guide 7 Service...
Page 94: ...80 Agilent 5517B BL C D DL FL User s Guide 8 Specifications...
Page 98: ......
Page 99: ......
Page 101: ......
Page 102: ...05517 90077 Manual Part Number 05517 90077 Printed in U S A JUNE 18 2009...